 Here you can set-up HV scans (current, noise, efficiency). Each scan would have a scan number on WEBDCS and each HV point would have a run number associated to it. In **Settings --> DCS Settings**. Be sure you choose Lyon DAQ. Also depending on the needs you may use **RPC Mode: Double gap or Single gap**. Here you can set-up HV scans (current, noise, efficiency). Each scan would have a scan number on WEBDCS and each HV point would have a run number associated to it. In **Settings --> DCS Settings**. Be sure you choose Lyon DAQ. Also depending on the needs you may use **RPC Mode: Double gap or Single gap**.
 +
 +Notice that here you would also set up the number of HV points.
  
 {{screenshot_from_2018-04-27_23_09_13.png}} {{screenshot_from_2018-04-27_23_09_13.png}}
  
 Then go to **HVscan --> Start DAQ HV scan**\\ Then go to **HVscan --> Start DAQ HV scan**\\
-Set up minimal measure time. It is the time for each HV point. It would be defined by the number of triggers expected with the current conditions. ​ASK EXPERTS ​how to get this time or get it from instructions ​in elog.\\+Set up minimal measure time. It is the time for each HV point. It would be defined by the number of triggers expected with the current conditions. ​To obtain it you can do a short measurement with an HV point and look on how many triggers you collect in few mn. 
 + 
 +To see it go to Monitor 6 on your left and type 
 + 
 +daqcontrol --daq-tdcstatus 
 + 
 +then look on how many triggers you collect ​in 1 spill. 
 + 
 +\\
 Keep waiting time at 5 mn.\\ Keep waiting time at 5 mn.\\
 Max Triggers have no impact on DAQ just have to be above 10.\\ Max Triggers have no impact on DAQ just have to be above 10.\\
Line 36: Line 46:
  
  
-Before you start the run you need to do some book keeping [[beamtest_gifpp ​| at the end of this page ]]. Prepare the table in following format and some explanation. \\+Before you start the run you need to do some book keeping [[tablesgif_042018 ​| at the end of the dokuwiki ​page containing all tables]]. Prepare the table in following format and some explanation. \\
 The HV points shall be the ones you prepare during "Start DAQ HV scan" phase. \\ The HV points shall be the ones you prepare during "Start DAQ HV scan" phase. \\
 The attenuator you know from the source console.\\ The attenuator you know from the source console.\\
Line 64: Line 74:
  
 __At the end of your shift please post a summary on elog__: [[https://​cmsonline.cern.ch/​webcenter/​portal/​cmsonline/​Common/​Elog?​__adfpwp_action_portlet=623564097&​__adfpwp_backurl=https%3A%2F%2Fcmsonline.cern.ch%3A443%2Fwebcenter%2Fportal%2Fcmsonline%2FCommon%2FElog%3FMedia-Type%3Dscreen%26Media-Feature-Scan%3D0%26Media-Feature-Orientation%3Dlandscape%26Media-Feature-Device-Height%3D1000%26Media-Feature-Height%3D808%26_afrWindowMode%3D0%26Media-Feature-Monochrome%3D0%26Font-Size%3D16%26Media-Feature-Color%3D8%26Media-Featured-Grid%3D0%26_afrLoop%3D32155132098559331%26Media-Feature-Resolution%3D115%26Media-Feature-Width%3D1333%26Media-Feature-Device-Width%3D1333%26Media-Feature-Color-Index%3D0%26Adf-Window-Id%3Dw0%26__adfpwp_mode.623564097%3D1&​_piref623564097.strutsAction=%2FviewSubcatMessages.do%3FcatId%3D891%26subId%3D893%26page%3D1 | elog ]] __At the end of your shift please post a summary on elog__: [[https://​cmsonline.cern.ch/​webcenter/​portal/​cmsonline/​Common/​Elog?​__adfpwp_action_portlet=623564097&​__adfpwp_backurl=https%3A%2F%2Fcmsonline.cern.ch%3A443%2Fwebcenter%2Fportal%2Fcmsonline%2FCommon%2FElog%3FMedia-Type%3Dscreen%26Media-Feature-Scan%3D0%26Media-Feature-Orientation%3Dlandscape%26Media-Feature-Device-Height%3D1000%26Media-Feature-Height%3D808%26_afrWindowMode%3D0%26Media-Feature-Monochrome%3D0%26Font-Size%3D16%26Media-Feature-Color%3D8%26Media-Featured-Grid%3D0%26_afrLoop%3D32155132098559331%26Media-Feature-Resolution%3D115%26Media-Feature-Width%3D1333%26Media-Feature-Device-Width%3D1333%26Media-Feature-Color-Index%3D0%26Adf-Window-Id%3Dw0%26__adfpwp_mode.623564097%3D1&​_piref623564097.strutsAction=%2FviewSubcatMessages.do%3FcatId%3D891%26subId%3D893%26page%3D1 | elog ]]
 +
 +
 +
 +==== **DAQ configuration for intensive source** ====
 +
 +To be done before source modification
 +
 +if source ATT > 46
 +
 +source att46_settings
 +daqcontrol --daq-setvth --vth=510
 +
 +look on top command.
  
  
Line 79: Line 102:
 **Run: DAQ run number to write in dokuwiki table**\\ **Run: DAQ run number to write in dokuwiki table**\\
  
 +You can also survey ​
  
 Possible problems: Possible problems:
Line 89: Line 112:
  
 [[expert_shifterinstructions_gifpp | The instructions for experts.]]\\ [[expert_shifterinstructions_gifpp | The instructions for experts.]]\\
 +
  
 ==== **DQM** ==== ==== **DQM** ====
  
 At some point some basic DQM would be implemented in WEBDCS. Till then you can [[gif_quick_tesk_042018 | run the analysis code described here]]. At some point some basic DQM would be implemented in WEBDCS. Till then you can [[gif_quick_tesk_042018 | run the analysis code described here]].
 +
 +
 +
  
  