This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision | ||
calibration_des_boards_fe_4_et_fe_2 [2017/11/27 11:12] acqilc [Resistance 150 kOhm] |
calibration_des_boards_fe_4_et_fe_2 [2021/07/10 23:21] (current) |
||
---|---|---|---|
Line 179: | Line 179: | ||
== Etude de l'efficacité == | == Etude de l'efficacité == | ||
+ | |||
+ | Seuil a **343** | ||
+ | |||
+ | **(343-327)/0.139 =115 fC /2 => 57.5 fC par canal** | ||
^ run ^ Injection(fC) ^ N trg ^ N 2 ^ N 14 ^ Eff ^ deFF ^ Eff XY ^ deff XY ^ $\Delta T $(ns) ^ $\sigma $ dt (ns) ^ | ^ run ^ Injection(fC) ^ N trg ^ N 2 ^ N 14 ^ Eff ^ deFF ^ Eff XY ^ deff XY ^ $\Delta T $(ns) ^ $\sigma $ dt (ns) ^ | ||
Line 314: | Line 318: | ||
</imgcaption> | </imgcaption> | ||
- | On met le seuil a 494 | + | |
+ | On met le seuil a **494** | ||
+ | |||
+ | **(494-471)/0.1845 = 125 fC /2 => 62.5 fC de seuil par canal** | ||
^ run ^ Injection(fC) ^ N trg ^ N 2 ^ N 14 ^ Eff ^ deFF ^ Eff XY ^ deff XY ^ dt (ns) ^ RMS dt (ns) ^ | ^ run ^ Injection(fC) ^ N trg ^ N 2 ^ N 14 ^ Eff ^ deFF ^ Eff XY ^ deff XY ^ dt (ns) ^ RMS dt (ns) ^ |