New firmware (v2) on both chambers

Mapping return chamber

Channel 0 is the TDC of trigger The strip 12 does not exist. in this mapping

For calibration you got TDC (channel -1) to Petiroc

      firmwaret=[31,29,27,25,23,22,21,20,19,18,17,16,15,14,13,12,11,10,9,8,7,6]

and TDC to strip

{"id":10,"chamber":1,"first":741000,"last":743000,
         "triggerMean":-180.,"triggerMin":-230.,"triggerMax":-150.,
         "stripShift":0,"polarity":1,
         "tdc2strip":[12,1,1,2,2,3,3,4,4,5,5,6,6,7,7,8,8,9,9,10,10,11,11,12],
         "side":[1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0],
         "delta":[0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0]
        },
        {"id":14,"chamber":1,"first":741000,"last":743000,
         "triggerMean":-180.,"triggerMin":-230.,"triggerMax":-150.,
         "stripShift":11,"polarity":-1,
         "tdc2strip":[12,11,11,10,10,9,9,8,8,7,7,6,6,5,5,4,4,3,3,2,2,1,1,12],
         "side":[1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0],
         "delta":[0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0]
        },
        {"id":11,"chamber":1,"first":741000,"last":743000,
         "triggerMean":-180.,"triggerMin":-230.,"triggerMax":-150.,
         "stripShift":22,"polarity":1,
         "tdc2strip":[12,1,1,2,2,3,3,4,4,5,5,6,6,7,7,8,8,9,9,10,10,11,11,12],
         "side":[1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0],
         "delta":[0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0]
        },
        {"id":9,"chamber":1,"first":741000,"last":743000,
         "triggerMean":-180.,"triggerMin":-230.,"triggerMax":-150.,
         "stripShift":33,"polarity":-1,
         "tdc2strip":[12,11,11,10,10,9,9,8,8,7,7,6,6,5,5,4,4,3,3,2,2,1,1,12],
         "side":[1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0,1,0],
         "delta":[0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0]
        }

Mapping coaxial chamber

Channel 0 is the TDC of trigger The strip 13 does not exist. in this mapping

For calibration you got TDC (channel -1) to Petiroc

      firmware2=[24,5,3,1,0,2,4,6,7,8,9,10,26,28,30,31,29,27,25,23,22,21,20,19]

and TDC to strip

{"id":5,"chamber":2,"first":741000,"last":743000,
         "triggerMean":-180.,"triggerMin":-230.,"triggerMax":-150.,
         "stripShift":0,"polarity":1,
         "tdc2strip":[13,1,2,3,4,5,6,7,8,9,10,11,12,1,2,3,4,5,6,7,8,9,10,11,12],
         "side":[2,1,1,1,1,1,1,1,1,1,1,1,1,0,0,0,0,0,0,0,0,0,0,0,0],
         "delta":[0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0]
        },
        {"id":6,"chamber":6,"first":741000,"last":743000,
         "triggerMean":-180.,"triggerMin":-230.,"triggerMax":-150.,
         "stripShift":12,"polarity":1,
          "tdc2strip":[13,1,2,3,4,5,6,7,8,9,10,11,12,1,2,3,4,5,6,7,8,9,10,11,12],
         "side":[2,1,1,1,1,1,1,1,1,1,1,1,1,0,0,0,0,0,0,0,0,0,0,0,0],
         "delta":[0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0]
        },
        {"id":7,"chamber":2,"first":741000,"last":743000,
         "triggerMean":-180.,"triggerMin":-230.,"triggerMax":-150.,
         "stripShift":22,"polarity":1,
         " "tdc2strip":[13,1,2,3,4,5,6,7,8,9,10,11,12,1,2,3,4,5,6,7,8,9,10,11,12],
         "side":[2,1,1,1,1,1,1,1,1,1,1,1,1,0,0,0,0,0,0,0,0,0,0,0,0],
         "delta":[0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0]
        },
        {"id":8,"chamber":2,"first":741000,"last":743000,
         "triggerMean":-180.,"triggerMin":-230.,"triggerMax":-150.,
         "stripShift":33,"polarity":-1,
          "tdc2strip":[13,1,2,3,4,5,6,7,8,9,10,11,12,1,2,3,4,5,6,7,8,9,10,11,12],
         "side":[2,1,1,1,1,1,1,1,1,1,1,1,1,0,0,0,0,0,0,0,0,0,0,0,0],
         "delta":[0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0, 0.0]
        }

Data format and software

The new data format is explain in this page

The detector id of the FEB is now 130

Example of readout of binary data can be found in this git directory] of my analysis. The //TdcChannel// class change (FEBCMS tag) to take into account the 6 bytes structure. The //read// method in tdcrb.cxx (FEBCMS tag) is addapted to this new format ===== Daq environnment ===== The configuration used is : **//all_firmware2_1//** <code> { "HOSTS": { "lyocmsrpi02": [ { "ARGS": [], "ENV": [ "PATH=/usr/local/sbin:/usr/local/bin:/usr/sbin:/usr/bin:/sbin:/bin", "DIM_DNS_NODE=lyosdhcal10", "LD_LIBRARY_PATH=/usr/lib:/usr/local/lib:/opt/dhcal/levbdim/lib:/opt/dhcal/lib:/opt/dhcal/DQM4HEP/lib:/opt/dhcal/dim/linux:/opt/dhcal/DQM4HEP/lib:/opt/dhcal/root/lib::/opt/dhcal/lcio/v02-00/lib/:$LD_LIBRARY_PATH", "WEBPORT=42000" ], "NAME": "MDCCSERVER", "PARAMETER": { "device": "MDCC01", "spilloff": 1000, "spillon": 1000000, "spillregister": 32 }, "PROGRAM": "/opt/dhcal/bin/lydaq_daemon_Mdcc" }, { "ARGS": [], "ENV": [ "LD_LIBRARY_PATH=/opt/dhcal/root/lib:/usr/lib:/usr/local/lib:/opt/dhcal/lib:$LD_LIBRARY_PATH", "WEBPORT=32000" ], "NAME": "GENESYS", "PARAMETER": { "TCPPort": 5600, "device": "/dev/ttyUSB0", "location": "CERN_904", "period": 30, "port": 6 }, "PROGRAM": "/opt/dhcal/bin/lydaq_daemon_Genesys" } ], "lyoilcrpi13": [ { "ARGS": [], "ENV": [ "LD_LIBRARY_PATH=/opt/dhcal/root/lib:/usr/lib:/usr/local/lib:/opt/dhcal/lib:$LD_LIBRARY_PATH", "WEBPORT=31000" ], "NAME": "HIH", "PARAMETER": { "TCPPort": 5600, "location": "CERN_904", "period": 30 }, "PROGRAM": "/opt/dhcal/bin/lydaq_daemon_hih8000" } ], "lyoilcrpi24": [ { "ARGS": [], "ENV": [ "LD_LIBRARY_PATH=/opt/dhcal/root/lib:/usr/lib:/usr/local/lib:/opt/dhcal/lib:$LD_LIBRARY_PATH", "WEBPORT=31000" ], "NAME": "BMP", "PARAMETER": { "TCPPort": 5600, "location": "CERN_904", "period": 30 }, "PROGRAM": "/opt/dhcal/bin/lydaq_daemon_Bmp" } ], "lyosdhcal11": [ { "ARGS": [], "ENV": [ "DIM_DNS_NODE=lyosdhcal11", "LD_LIBRARY_PATH=/usr/lib:/usr/local/lib:/opt/dhcal/levbdim/lib:/opt/dhcal/lib:/opt/dhcal/root/lib:/opt/dhcal/dim/linux:/opt/dhcal/lcio/v02-00/lib/:$LD_LIBRARY_PATH", "WEBPORT=45000" ], "NAME": "FDAQ", "PARAMETER": { "s_ctrlreg": "0x815A1B40" }, "PROGRAM": "/opt/dhcal/bin/lydaq_daemon_Control" }, { "ARGS": [], "ENV": [ "PATH=/usr/local/sbin:/usr/local/bin:/usr/sbin:/usr/bin:/sbin:/bin", "LD_LIBRARY_PATH=/usr/lib:/usr/local/lib:/opt/dhcal/levbdim/lib:/opt/dhcal/lib:/opt/dhcal/DQM4HEP/lib:/opt/dhcal/dim/linux:/opt/dhcal/DQM4HEP/lib:/opt/dhcal/root/lib::/opt/dhcal/lcio/v02-00/lib/:$LD_LIBRARY_PATH", "WEBPORT=40000" ], "NAME": "TDCSERVER", "PARAMETER": { "publish": "tcp://lyosdhcal11:5556", "tdc": { "db": { "mode": "DB", "state": "TDC_RETNEW_9" }, "network": "192.168.10." } }, "PROGRAM": "/opt/dhcal/bin/lydaq_daemon_FebCms" }, { "ARGS": [], "ENV": [ "LD_LIBRARY_PATH=/usr/lib:/usr/local/lib:/opt/dhcal/lib:/opt/dhcal/lcio/v02-00/lib/:/opt/dhcal/root/lib:$LD_LIBRARY_PATH", "WEBPORT=51000" ], "NAME": "WRITER", "PARAMETER": { "dif": 1, "directory": "/data/local/FE1PR2", "processor": [ "binarywriter", "lydaq_pluggins_monitor" ], "shmpath": "/dev/shm/monitor", "stream": [ "tcp://*:5556" ] }, "PROGRAM": "/opt/dhcal/bin/lydaq_daemon_Builder" }, { "ARGS": [], "ENV": [ "LD_LIBRARY_PATH=/usr/lib:/usr/local/lib:/opt/dhcal/lib:$LD_LIBRARY_PATH", "WEBPORT=36000" ], "NAME": "FSLOW", "PARAMETER": { "account": "admin/admin@lyoac28", "deviceName": "SY1527@DOMESTAGEM2", "first": 0, "last": 5, "serverName": "tcp://*:5500" }, "PROGRAM": "/opt/dhcal/bin/lydaq_daemon_Slow_Control" } ] }, "NAME": "all_firmware2_1" } </code> It requires 4 computers: * lyosdhcal11: where all FEB readout, Event building and DAQ control run * lyocmsrpi02: It controls the MDCC and the Genesys Low Volatge power supply * lyoilcrpi13: It measures the humidity and temperature on 2 HIH8000 probes * lyoilcrpi24: It measures temperature and Presssure with a BMP183 probe Data are stored on ''/data/local/FE1PR2'' ==== ASICs configuration ==== The initial configuration with both the return chamber calibrated (TDC_RETNEW_8) and the 4 FEBs of the coaxial chamber is **TDC_RETNEW_9** Two runs were use to calibrate those Febs to 480 VTHTIME: 741263 and 741278 Nevertheless the HV was set to 7800 V when it was done and detector noise pollutes those runs. The new state to be used for cosmic data taking is **TDC_RETNEW_12** at 1 kV on all chambers we got run 741284 {{ :cern_july18:run741284_allstrip5.png |}} {{ :cern_july18:run741284_allstrip6.png |}} {{ :cern_july18:run741284_allstrip7.png |}} {{ :cern_july18:run741284_allstrip8.png |}} The return strip chamber is a bit noisy (ground connection to be adjusted) and **the minimal threshold is 504** ( ~ 65 fC) --- //[[laurent.mirabito@gmail.com|DAQ user 2018/07/05 13:53 <pre> acqilc@lyosdhcal11:~$ slccontrol –slc-ptstatus P T 962.00 295.45 acqilc@lyosdhcal11:~$ slccontrol –slc-humstatus H0 T0 H1 T1 46.53 294.71 37.14 294.89 </pre>